Structural optimization of orthopedic hip implant using parametric and non-parametric optimization techniques

Biomed Phys Eng Express. 2023 Aug 16;9(5). doi: 10.1088/2057-1976/aced0d.

Abstract

This research investigates the reduction in weight of hip implant by the application of parametric and non-parametric optimization techniques. Orthopaedic hip implants can be made from metals, ceramics, composites, or metallic alloys and are generally solid structures. The stiffness of orthopaedic hip implant is a pertaining problem when implanted in the human body as Hip implant are stiffer than bone material and causes stress shielding. This results in bone weakening which causes osteoporosis. Reduction in mass of femur stem results in stiffness reduction of femur stem. Non-Parametric topology optimization results in 34.9% mass reduction and parametric optimization based on Central Composite Design technique in Design of Experiments (DoE) uses hole diameters as parameters and performs structural optimization that results in 22% mass reduction.

Keywords: central composite design; design of experiments; finite element analysis; orthopedic hip implant; structural optimisation; topology optimisation.

MeSH terms

  • Alloys
  • Arthroplasty, Replacement, Hip* / methods
  • Hip Prosthesis*
  • Humans
  • Orthopedics*
  • Prosthesis Design

Substances

  • Alloys