Interaction between maintenance variables of medical ultrasound scanners through multifactor dimensionality reduction

Expert Rev Med Devices. 2023 Jul-Dec;20(10):851-864. doi: 10.1080/17434440.2023.2243208. Epub 2023 Aug 7.

Abstract

Background: Proper maintenance of electro-medical devices is crucial for the quality of care to patients and the economic performance of healthcare organizations. This research aims to identify the interaction between Ultrasound scanners (US) maintenance variables as a function of maintenance indicators: US in service or decommissioned, excessive number of failures, and failure rate. Knowing those interactions, specific maintenance measures will be developed to improve the reliability of the US.

Research design and methods: Multifactor Dimensionality Reduction (MDR) method was eployed to analyze data from 222 US and their four-year maintenance history. Models were developed based on the variables with the greatest influence on maintenance indicators, where US were classified according to the associated risk.

Results: US with more than one major failure or at least one major component replacement had up to 496.4% more failures than the average. Failure rate increased by up to 188.7% over the average for those US with more than three moderate failures, three replacements, or both.

Conclusions: This study identifies and quantifies the causes of risk to establish a specific maintenance plan for US. It helps to better understand the degradation of US to optimize their operation and maintenance.

Keywords: Electro-medical devices; healthcare engineering; maintenance; multifactor dimensionality reduction; reliability; ultrasound scanner.