Synchronization of light flash with the irradiation pulse in proton beam therapy: A case report

Tech Innov Patient Support Radiat Oncol. 2023 Jul 7:27:100218. doi: 10.1016/j.tipsro.2023.100218. eCollection 2023 Sep.

Abstract

The correlation between sensory light flash and proton beam delivery was evaluated by measuring the timing of pulse beam delivery and light flash sensing using an event recorder in an 83-year-old patient receiving proton beam therapy (PBT) for nasopharyngeal adenoid cystic carcinoma. The treatment dose was 65 Gy (RBE) in 26 fractions with 2 ports, and both beams included the visual pathway (retina, optic nerve, chiasma). Measurements were obtained in 13 of the 26 fractions. The patient sensed a light flash in all 13 fractions and pressed the recorder button for 426 of the 430 pulsed beam deliveries, giving a sensing rate of 99.1%. The median duration of button-pressing of 0.3 s was almost the same as that of the beam pulse of 0.2 s, with a reaction time lag of 0.35 s. These results suggest a consistency between light flash during PBT and the timing of irradiation.

Keywords: Irradiation pulse; Light flash; Proton beam therapy; Quantitative.

Publication types

  • Case Reports