Development and multi-institutional evaluation of a new phantom for verifying beam-positioning errors at off-isocenter positions

Phys Med. 2023 Aug:112:102645. doi: 10.1016/j.ejmp.2023.102645. Epub 2023 Jul 19.

Abstract

Purpose: Single-isocenter stereotactic radiotherapy for multiple brain metastases requires highly accurate treatment delivery at off-isocenter positions (off-iso). This study aimed to verify the beam-positioning errors at off-iso using a newly developed phantom tested at multiple institutions.

Methods: The off-iso phantom comprised five stainless-steel balls with a 3-mm diameter placed at the center and at four peripheral positions on a diagonal line. Each ball was placed 3.5 cm apart along each of the three axes. Two patterns of the phantom setup were defined as 0° and 90° phantom rotations to evaluate the beam-positioning error, which is the distance between the center of the ball and the irradiated field on the electronic portal imaging device. Furthermore, the reproducibility of the beam-positioning errors was verified by evaluating their standard deviation (SD) at a single institution, which included five measurements for two treatment machines. The errors were evaluated at multiple institutions using eight treatment machines.

Results: The measurement time from setup to image acquisition was approximately 20 min for two patterns. The SD of the beam-positioning errors in the reproducibility tests was 0.41 mm. In the multi-institutional evaluation, the beam-positioning error at the isocenter position was within 1.00 mm of the AAPM-RSS tolerance, with the exception of two linacs. The largest beam-positioning error (1.36 mm) was observed 7.5 cm away from the isocenter in three directions at a gantry angle of 180°.

Conclusions: The developed phantom can be applied as a new tool for establishing beam-positioning errors in single-isocenter stereotactic radiotherapy at off-isocenter positions.

Keywords: Beam-positioning error; Multi-institutional evaluation; Off isocenter; Single-isocenter stereotactic radiotherapy.

Publication types

  • Multicenter Study

MeSH terms

  • Brain Neoplasms* / radiotherapy
  • Humans
  • Phantoms, Imaging
  • Radiosurgery* / methods
  • Radiotherapy Planning, Computer-Assisted / methods
  • Reproducibility of Results