A quantitative comparison of devices for in vivo biomechanical characterization of human skin

Mech Soft Mater. 2023;5(1):5. doi: 10.1007/s42558-023-00053-w. Epub 2023 Jul 17.

Abstract

Non-invasive skin characterization devices are emerging as a valuable tool in clinical skin research. In recent years, the range of available experimental techniques and methods used to determine the biomechanical properties of skin has increased considerably. Although a substantial amount of work has been devoted to assessing the working principle of macroscopic skin characterization devices individually, a rationalization and comparison between them is still lacking. This motivated the present study, which aimed to characterize and compare three commonly used working principles: suction, dynamic shear loading, and indentation. A synthetic model system with tunable mechanical properties was used to assess the three devices, and the results rationalized based on corresponding finite element models. In vivo measurements were performed on healthy volunteers to investigate the capability of differentiating the biomechanical properties of skin at different body locations, and to assess the intra- and inter-rater reliability of each device. The present comparative analysis indicates that the analyzed functional principles perceive the stiffness of human skin differently, with relevant implications for the interpretation of the respective measurement results.

Keywords: Myoton; Nimble; Skin biomechanics; SkinFibroMeter.