A quantitative description of the compatibility of technetium-selective chromatographic technetium-99m separation with low specific activity molybdenum-99

J Chromatogr A. 2023 Aug 30:1705:464192. doi: 10.1016/j.chroma.2023.464192. Epub 2023 Jul 3.

Abstract

Technetium-99m generators employing a technetium-selective stationary phase are a chromatographic instrument developed for use with 99Mo having low specific activity (LSA); particularly, 99Mo produced by electron accelerators. This paper presents a mathematical description of technetium-selective chromatographic (TSC) 99mTc separation and analyzes its compatibility with LSA 99Mo. We developed a theoretical formula for TSC 99mTc separation by discretizing its pertechnetate selectivity, and validated it using an electron linear accelerator and activated carbon-based TSC (AC-TSC) 99mTc generators. We confirmed that the activity concentration of 99mTc obtained from a TSC 99mTc generator can be calculated directly from its input 99Mo activity regardless of the 99Mo specific activity. The formula corroborates that TSC 99mTc separation is compatible with LSA 99Mo, and has a practical application in estimating the number of TSC 99mTc generators required for 99mTc demand of interest.

Keywords: Activity concentration; Chromatographic separation; Molybdenum-99; Specific activity; Technetium-99m.

MeSH terms

  • Electrons
  • Molybdenum / chemistry
  • Radioisotopes*
  • Technetium* / chemistry

Substances

  • Molybdenum-99
  • Technetium
  • Radioisotopes
  • Molybdenum