In vivo creep and wear performance of vitamin-E-diffused highly crosslinked polyethylene in total hip arthroplasty

Arch Orthop Trauma Surg. 2023 Dec;143(12):7195-7203. doi: 10.1007/s00402-023-04972-1. Epub 2023 Jul 12.

Abstract

Introduction: An acetabular liner thickness of around 6 mm remains the "gold standard" in total hip arthroplasty. Some surgeons have been recommending the use of the thickest possible liner because contact stress and strain in articulating surfaces decrease with increasing the wall thickness. The purpose of this study was to determine whether in vivo creep and wear performance could be enhanced using a thicker liner over the standard thickness in vitamin-E-diffused highly crosslinked polyethylene (HXLPE).

Materials and methods: One hundred and twenty-two hips were allocated to age-matched, sex-matched, and body mass index-matched two subgroups implanted either with a 6.8- or 8.9-mm-thick vitamin-E-diffused HXLPE liner against 28-mm cobalt-chrome femoral head, and followed-up for 7 years. Linear and volumetric penetration of femoral head into the liners attributed to creep and wear were analyzed for each group.

Results: Compressive creep strain generated at the initial 6 months was significantly larger in the 6.8-mm group (2.6%) than in the 8.9-mm group (2.2%). The linear steady-state wear observed after 2 years was 0.0019 and 0.0015 mm/year, whereas the volumetric steady-state wear was 0.54 and 0.45 mm3/years in the 6.8- and 8.9-mm-thick groups, respectively. Although less strain in the thicker group resulted in a slightly less wear, it did not reach significant differences in the steady-state wear rates between the groups.

Conclusion: No clinical significance for using a thicker liner over the standard thickness (6.8 mm → 8.9 mm) was confirmed in the vitamin-E-diffused HXLPE according to the 7-year follow-up. The wear rates for both thicknesses were very low enough to prevent osteolysis, and no mechanical failure was observed at any follow-up interval. Nevertheless, since the significantly higher strain was seen in the thinner liner, further follow-up is needed to compare the longer term wear and the incidence of osteolysis and component fracture.

Keywords: Creep; Highly crosslinked polyethylene; Total hip arthroplasty; Vitamin E; Wear.

MeSH terms

  • Arthroplasty, Replacement, Hip* / adverse effects
  • Child
  • Follow-Up Studies
  • Hip Prosthesis* / adverse effects
  • Humans
  • Osteolysis* / etiology
  • Polyethylene
  • Prosthesis Design
  • Prosthesis Failure
  • Vitamin E
  • Vitamins

Substances

  • Polyethylene
  • Vitamin E
  • Vitamins