Accuracy of Microimplant Placement Using a 3D Guide Plate for Orthodontic Anchorage

Appl Bionics Biomech. 2023 Jun 26:2023:9060046. doi: 10.1155/2023/9060046. eCollection 2023.

Abstract

This study aimed to design a three-dimensional (3D) guide plate using computer-aided design and a 3D printing system for precise implantation of microimplants for orthodontic treatment and investigate the accuracy and feasibility of a 3D guide plate in clinical practice. A total of 30 microimplants were placed in 15 patients in the Department of Stomatology, Affiliated Hospital of Jiangnan University. Before surgery, DICOM data from cone-beam computed tomography (CBCT) scans and STereoLithography data from the 3D model scan were imported to 3Shape Dental System. Data fitting and matching were performed, and 3D guide plates were designed primarily focusing on the thickness of guide plates, amount of concave compensation, and dimensions of the ring. Assist implantation method was used to place the microimplants, and postoperative CBCT images were used to evaluate the position and implantation angle. The feasibility of placing microimplants and precise implantation guided by the 3D guide plate. CBCT data before and after the placement of microimplants were compared. Regarding the secure positioning of microimplants based on CBCT data, 26 implants were categorized as Grade i, four as Grade ii, and none as Grade iii. No loosening of microimplants 1 and 3 months after surgery was reported. The implantation of microimplants is more accurate under the guidance of a 3D guide plate. This technology can achieve accurate implant positioning, thus ensuring safety, stability, and improved success rates after implantation.