Persistent flow inside the Woven EndoBridge at angiographic follow-up: A multicenter study

Interv Neuroradiol. 2023 Jul 4:15910199231185805. doi: 10.1177/15910199231185805. Online ahead of print.

Abstract

Background: The occurrence of persistent intra-device filling (BOSS 1, using the Bicêtre Occlusion Scale Score (BOSS)) in aneurysms treated with a Woven Endobridge (WEB) device is infrequent based on angiographic follow-up. To date, three monocentric case series were published studying BOSS 1 cases. Through a multicenter retrospective study, we aimed to report the incidence, and risk factors of intra-WEB persistent filling.

Methods: We reached out to European academic centers that treat patients using WEB devices and requested de-identified data of patients treated with a WEB device and underwent angiographic follow-up, at least 3 months after embolization, to assess the BOSS 1 occlusion score. We compared baseline characteristics, treatment modalities, and aneurysm data of the included BOSS 1 patients with those of a control group consisting of non-BOSS 1 patients (n = 116) who had an available angiographic follow-up. Univariable and multivariable models were employed for analysis.

Results: Among the pooled sample of 591 aneurysms treated with WEB, the rate of persistent flow (BOSS 1) at angiographic follow-up was 5.2% (n = 31 out of 591), performed after an average of 8.7 ± 6.3 months. In the multivariable-adjusted analysis, dual antiplatelet therapy in the postoperative period (adjusted odds ratio [aOR] 4.3 [95% CI 1.3-14.2]), and WEB undersizing (aOR 10.8 [95% CI 2.9-40]) were independently associated with a BOSS 1 persistent flow result.

Conclusion: Persistent blood flow within the WEB device during angiographic follow-up (BOSS 1) is an uncommon occurrence. Our findings indicate that post-procedural dual antiplatelet therapy and undersizing of the WEB device are independently associated with the presence of BOSS 1 at follow-up.

Keywords: Aneurysm; Woven Endobridge; flow-disruption.