Aim: This study evaluated and compared the shaping ability of four rotary instrument systems in long-oval root canals using microcomputed tomographic (micro-CT) evaluation. Currently, there is no data available on the canal shaping abilities of BlueShaper and DC Taper instruments.
Methodology: Sixty-four single-rooted mandibular premolars were matched based on similar root canal morphologic features as determined by (micro-CT) and randomly assigned to 1 of 4 experimental groups (n = 16) according to the instrument system used: BlueShaper, TruNatomy, DC Taper and HyFlex EDM One File. Changes in the root canal surface and volume, remaining dentine thickness, and number of prepared areas were assessed.
Results: No significant differences were found amongst the four instrument systems for the parameters evaluated (p > .05). There was a significant reduction in the number of unprepared areas and the remaining dentine thickness after each increase in size of the instruments tested (p < .05).
Conclusions: The four instrument systems perform similarly in long oval root canals. Although none could prepare all canal walls, larger preparations incorporated significantly more surfaces in the final shape.
Keywords: micro-computed tomography; oval root canals; root canal preparation; shaping ability.