An accelerator-based boron neutron capture therapy (AB-BNCT) system was installed at the Shonan Kamakura General Hospital (SKGH). We confirmed that a stable operation was possible for 1 h at a current of 30 mA. The evaluated thermal neutron flux was 2.8 × 109 cm-2 s-1 and in good agreement (±5%) with the calculated values. The daily variation was within ±2%. The ambient dose rate due to residual radioactivity after irradiation was approximately 5 μSv/h using a lead shutter.
Keywords: Accelerator-based neutron source; Boron neutron capture therapy; Cockcroft-walton type electrostatic accelerator; Quality assurance.