Initial evaluation of accelerator-based neutron source system at the Shonan Kamakura General Hospital

Appl Radiat Isot. 2023 Sep:199:110898. doi: 10.1016/j.apradiso.2023.110898. Epub 2023 Jun 8.

Abstract

An accelerator-based boron neutron capture therapy (AB-BNCT) system was installed at the Shonan Kamakura General Hospital (SKGH). We confirmed that a stable operation was possible for 1 h at a current of 30 mA. The evaluated thermal neutron flux was 2.8 × 109 cm-2 s-1 and in good agreement (±5%) with the calculated values. The daily variation was within ±2%. The ambient dose rate due to residual radioactivity after irradiation was approximately 5 μSv/h using a lead shutter.

Keywords: Accelerator-based neutron source; Boron neutron capture therapy; Cockcroft-walton type electrostatic accelerator; Quality assurance.

MeSH terms

  • Boron Neutron Capture Therapy* / methods
  • Hospitals, General*
  • Neutrons