Optimisation and implementation of QuEChERS-based sample preparation for identification and semi-quantification of 694 targeted contaminants in honey, jam, jelly, and syrup by UHPLC-Q/ToF high-resolution mass spectrometry

Food Chem. 2023 Nov 1:425:136448. doi: 10.1016/j.foodchem.2023.136448. Epub 2023 May 24.

Abstract

A screening and semi-quantitative method was developed for the analysis of 694 various contaminants in honey, jam, jelly and syrup samples by ultrahigh-performance liquid chromatography and quadrupole time-of-flight mass spectrometry. Sample preparation, which was optimised using split factorial design, was based on acetate-buffered version of QuEChERS, followed by a clean-up step and a concentration step to enhance sensitivity of analytes. The method was validated according to SANTE/11312/2021 guidelines. The screening detection and limits of identification were established as being less than or equal to 0.05 mg.kg-1 for 89% and 74% of the contaminants, respectively. The validated screening method was applied to 50 concentrated sugary products. Overall, 46% of the samples were positive to pesticide residues. Most of the positive samples (78%) contained mixtures of pesticide residues. Three time-and-cost saving convenient strategies suitable for high-throughput analysis were proposed for the targeted semi-quantification of the previously contaminants identified in samples.

Keywords: Concentrated sugary products; High-resolution mass spectrometry; Method validation; Pesticide residues; Semi-quantification.

MeSH terms

  • Chromatography, High Pressure Liquid / methods
  • Chromatography, Liquid
  • Honey* / analysis
  • Mass Spectrometry / methods
  • Pesticide Residues* / analysis

Substances

  • Pesticide Residues