Intracranial aneurysms treated with stent-assisted coil embolization: evaluation with four-dimensional ultrashort-TE MR angiography

Eur Radiol. 2023 Nov;33(11):7923-7933. doi: 10.1007/s00330-023-09755-1. Epub 2023 Jun 7.

Abstract

Objectives: As a novel follow-up method for intracranial aneurysms treated with stent-assisted coil embolization (SACE), we developed four-dimensional magnetic resonance angiography (MRA) with minimized acoustic noise utilizing ultrashort-echo time (4D mUTE-MRA). We aimed to assess whether 4D mUTE-MRA is useful for the evaluation of intracranial aneurysms treated with SACE.

Methods: This study included 31 consecutive patients with intracranial aneurysm treated with SACE who underwent 4D mUTE-MRA at 3 T and digital subtraction angiography (DSA). For 4D mUTE-MRA, five dynamic MRA images with a spatial resolution of 0.5 × 0.5 × 0.5 mm3 were obtained every 200 ms. Two readers independently reviewed the 4D mUTE-MRA images to evaluate the aneurysm occlusion status (total occlusion, residual neck, and residual aneurysm) and the flow in the stent using a 4-point scale (from 1 [not visible] to 4 [excellent]). The interobserver and intermodality agreement was assessed using κ statistics.

Results: On DSA images, 10 aneurysms were classified as total occlusion, 14 as residual neck, and 7 as residual aneurysm. In terms of aneurysm occlusion status, the intermodality and interobserver agreement was excellent (κ = 0.92 and κ = 0.96, respectively). For the flow in the stents on 4D mUTE-MRA, the mean score was significantly higher for single stents than multiple stents (p < .001) and for open-cell type stents than closed-cell type (p < .01).

Conclusions: 4D mUTE-MRA is a useful tool with a high spatial and temporal resolution for the evaluation of intracranial aneurysms treated with SACE.

Key points: • In the evaluation of intracranial aneurysms treated with SACE on 4D mUTE-MRA and DSA, the intermodality and interobserver agreement in aneurysm occlusion status was excellent. • 4D mUTE-MRA shows good to excellent visualization of flow in the stents, especially for cases treated with a single or open-cell stent. • 4D mUTE-MRA can provide hemodynamic information related to embolized aneurysms and the distal arteries to stented parent arteries.

Keywords: Intracranial aneurysm; Magnetic resonance angiography; Magnetic resonance imaging; Stents.

MeSH terms

  • Angiography, Digital Subtraction / methods
  • Embolization, Therapeutic* / methods
  • Follow-Up Studies
  • Humans
  • Intracranial Aneurysm* / diagnostic imaging
  • Intracranial Aneurysm* / therapy
  • Magnetic Resonance Angiography / methods
  • Stents
  • Treatment Outcome