Bias and Precision of Continuous P0.1 Measurement by Various Ventilators: A Simulation Study

Respir Care. 2023 Oct;68(10):1393-1399. doi: 10.4187/respcare.10755. Epub 2023 May 23.

Abstract

Background: Most ventilators measure airway occlusion pressure (occlusion P0.1) by occluding the breathing circuit; however, some ventilators can predict P0.1 for each breath without occlusion. Nevertheless, few studies have verified the accuracy of continuous P0.1 measurement. The aim of this study was to evaluate the accuracy of continuous P0.1 measurement compared with that of occlusion methods for various ventilators using a lung simulator.

Methods: A total of 42 breathing patterns were validated using a lung simulator in combination with 7 different inspiratory muscular pressures and 3 different rise rates to simulate normal and obstructed lungs. PB980 and Dräger V500 ventilators were used to obtain occlusion P0.1 measurements. The occlusion maneuver was performed on the ventilator, and a corresponding reference P0.1 was recorded from the ASL5000 breathing simulator simultaneously. Hamilton-C6, Hamilton-G5, and Servo-U ventilators were used to obtain sustained P0.1 measurements (continuous P0.1). The reference P0.1 measured with the simulator was analyzed by using a Bland-Altman plot.

Results: The 2 lung mechanical models capable of measuring occlusion P0.1 yielded values equivalent to reference P0.1 (bias and precision values were 0.51 and 1.06, respectively, for the Dräger V500, and were 0.54 and 0.91, respectively, for the PB980). Continuous P0.1 for the Hamilton-C6 was underestimated in both the normal and obstructive models (bias and precision values were -2.13 and 1.91, respectively), whereas continuous P0.1 for the Servo-U was underestimated only in the obstructive model (bias and precision values were -0.86 and 1.76, respectively). Continuous P0.1 for the Hamilton-G5 was mostly similar to but less accurate than occlusion P0.1 (bias and precision values were 1.62 and 2.06, respectively).

Conclusions: The accuracy of continuous P0.1 measurements varies based on the characteristics of the ventilator and should be interpreted by considering the characteristics of each system. Moreover, measurements obtained with an occluded circuit could be desirable for determining the true P0.1.

Keywords: P0.1; respiratory drive; airway occlusion pressure; bench test; lung simulator; mechanical ventilation; ventilator circuit; work of breathing.

MeSH terms

  • Computer Simulation
  • Equipment Design
  • Humans
  • Lung
  • Respiration, Artificial*
  • Ventilators, Mechanical*