Predicting Inhaled Drug Dose Generated by Mesh Nebulizers

J Aerosol Med Pulm Drug Deliv. 2023 Aug;36(4):162-170. doi: 10.1089/jamp.2022.0055. Epub 2023 May 19.

Abstract

Background: The lung dose of nebulized drugs for spontaneous breathing is influenced by breathing patterns and nebulizer performance. This study aimed to develop a system for measuring breath patterns and a formula for estimating inhaled drugs, and then to validate the hypothesized prediction formula. Methods: An in vitro model was first used to determine correlations among the delivered dose, breath patterns, and doses deposited on the accessories and reservoirs testing with a breathing simulator to generate 12 adult breathing patterns (n = 5). A pressure sensor was developed to measure breathing parameters and used along with a prediction formula that accounted for the initial charge dose, respiratory pattern, and dose on the accessory and reservoir of a nebulizer. Three brands of nebulizers were tested by placing salbutamol (5.0 mg/2.5 mL) in the drug holding chamber. Ten healthy individuals participated in the ex vivo study to validate the prediction formula. The agreement between the predicted and inhaled doses was analyzed using the Bland-Altman plot. Results: The in vitro model showed that the inspiratory time to total respiratory cycle time (Ti/Ttotal; %) was significantly directly correlated with the delivered dose among the respiratory factors, followed by inspiratory flow, respiratory rate, and tidal volume. The ex vivo model showed that Ti/Ttotal was significantly directly correlated with the delivered dose among the respiratory factors, in addition to the nebulization time and accessory dose. The Bland-Altman plots for the ex vivo model showed similar results between the two methods. Large differences in inhaled dose measured at the mouth were observed among the subjects, ranging from 12.68% to 21.68%; however, the difference between the predicted dose and inhaled dose was lower, at 3.98%-5.02%. Conclusions: The inhaled drug dose could be predicted with the hypothesized estimation formula, which was validated by the agreement between the inhaled and predicted doses of breathing patterns of healthy individuals.

Keywords: breath pattern; inhale drug dose; lung models; prediction formula; vibrating mesh nebulizers.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Administration, Inhalation
  • Adult
  • Aerosols
  • Albuterol
  • Bronchodilator Agents*
  • Equipment Design
  • Humans
  • Nebulizers and Vaporizers*

Substances

  • Aerosols
  • Albuterol
  • Bronchodilator Agents