Computer-Aided Detection of Subsolid Nodules on Chest Computed Tomography: Assessment of Visualization on Vessel-Suppressed Images

J Comput Assist Tomogr. 2023 May-Jun;47(3):412-417. doi: 10.1097/RCT.0000000000001444. Epub 2023 Mar 3.

Abstract

Objectives: This study aimed to clarify the performance of automatic detection of subsolid nodules by commercially available software on computed tomography (CT) images of various slice thicknesses and compare it with visualization on the accompanying vessel-suppression CT (VS-CT) images.

Methods: A total of 95 subsolid nodules from 84 CT examinations of 84 patients were included. The reconstructed CT image series of each case with 3-, 2-, and 1-mm slice thicknesses were loaded into a commercially available software application (ClearRead CT) for automatic detection of subsolid nodules and generation of VS-CT images. Automatic nodule detection sensitivity was assessed for 95 nodules on each series of images acquired at 3 slice thicknesses. Four radiologists subjectively evaluated visual assessment of the nodules on VS-CT.

Results: ClearRead CT automatically detected 69.5% (66/95 nodules), 68.4% (65/95 nodules), and 70.5% (67/95 nodules) of all subsolid nodules in 3-, 2-, and 1-mm slices, respectively. The detection rate was higher for part-solid nodules than for pure ground-glass nodules at all slice thicknesses. In the visualization assessment on VS-CT, 3 nodules at each slice thickness (3.2%) were judged as invisible, while 26 of 29 (89.7%), 27 of 30 (90.0%), and 25 of 28 (89.3%) nodules, which were missed by computer-aided detection, were judged as visible in 3-, 2-, and 1-mm slices, respectively.

Conclusions: The automatic detection rate of subsolid nodules by ClearRead CT was approximately 70% at all slice thicknesses. More than 95% of subsolid nodules were visualized on VS-CT, including nodules undetected by the automated software. Computed tomography acquisition at slices thinner than 3 mm did not confer any benefits.

MeSH terms

  • Computers
  • Humans
  • Lung Neoplasms*
  • Sensitivity and Specificity
  • Software
  • Tomography, X-Ray Computed / methods