Advances in Design and Development of Lumi-Solve: A Novel Drug-Eluting Photo-Angioplasty Device

Cardiovasc Eng Technol. 2023 Aug;14(4):605-614. doi: 10.1007/s13239-023-00668-0. Epub 2023 May 10.

Abstract

Purpose: The Lumi-Solve photo-angioplasty drug eluting balloon catheter (DEBc) may afford safety advantages over current DEBc. Lumi-Solve utilises the guidewire (GW) port and lumen to deliver fibre-optic UV365nm light to the angioplasty balloon which may be problematic. We explore and evaluate alternative Lumi-Solve design options to circumvent fibre-optic use of the GW port and lumen which may enhance efficacy and clinical utility.

Methods: Effects of guidewire shadowing (GWS) on visible and UV365nm light transmission were evaluated and modelled in-silico. To evaluate the effect of a dedicated intra-balloon fibre-optic port, modified angioplasty balloons and sections of translucent polyethylene terephthalate (PET) GW port tubing were utilised. Investigation of the effect of GWS on chemical and biological photo-activation of balloon surface drug was performed utilising LCMS analysis and inhibition of histone deacetylase activity (HDACi) was measured in human umbilical vein endothelial cells (HUVEC).

Results: Parallel fibre-optic and GW port configurations generated a GWS of approximately 18.0% of the evaluable balloon surface area and attenuated both visible and UV light intensity by 20.0-25.0% and reduced chemical photo-activation of balloon surface drug and HDACi by at least 40-45%. Alternative fibre-optic port configurations including a spiral design significantly mitigated GWS effects on UV light transmission.

Conclusions: To avoid use of the GW port and its associated complications a dedicated third port and lumen for the Lumi-Solve fibre-optic may be required. To maximize balloon surface chemical and biological photo-activation, non-parallel, intra-balloon, fibre-optic lumen trajectories, including a spiral design may be useful.

Keywords: Drug eluting balloon catheter; Fibre-optic; Guide wire shadow; Neointimal hyperplasia; Photo-activation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Angioplasty, Balloon* / adverse effects
  • Human Umbilical Vein Endothelial Cells
  • Humans
  • Vascular Access Devices*