Research on fatigue optimization simulation of polymeric heart valve based on the iterative sub-regional thickened method

Int J Numer Method Biomed Eng. 2023 Oct;39(10):e3717. doi: 10.1002/cnm.3717. Epub 2023 May 9.

Abstract

Prosthetic polymeric heart valves (PHVs) have the potential to overcome the inherent material and design limitations of traditional valves in the treatment of valvular heart disease; however, their durability remains limited. Optimal design of the valve structure is necessary to improve their durability. This study aimed to enhance the fatigue resistance of PHVs by improving the stress distribution. Iterative subregional thickening of the leaflets was used, and the mechanical stress distribution and hemodynamics of these polymeric tri-leaflet valves were characterized using a fluid-structure interaction approach. Subregional thickening led to a reduction in stress concentration on the leaflet, with the effective orifice area still meeting ISO 5840-3 and the regurgitant volume achieving a similar value to those in previous studies. The maximum stress in the final iteration was reduced by 28% compared with that of the prototype. The proposed method shows potential for analyzing the stress distribution and hemodynamic performance of subregional thickened valves and can further improve the durability of PHVs.

Keywords: Fluid-structure interaction (FSI); fatigue optimization; iterative; polymeric heart valve; subregional thickened.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aortic Valve
  • Computer Simulation
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Models, Cardiovascular
  • Polymers
  • Prosthesis Design
  • Stress, Mechanical

Substances

  • Polymers