Formation of spread-out Bragg peak for helium-ion beam using microdosimetric kinetic model

Phys Med. 2023 May:109:102587. doi: 10.1016/j.ejmp.2023.102587. Epub 2023 Apr 21.

Abstract

Purpose: To evaluate the applicability of microdosimetric kinetic model (MKM) to helium-ion therapy by forming a spread-out Bragg peak (SOBP) of a helium-ion beam using the MKM developed for carbon-ion radiotherapy and confirming the predictions in biological experiments.

Methods: Using a ridge filter, a 90-mm wide SOBP for a 210 MeV/u helium-ion beam was created in a broad beam delivery system. The ridge filter was designed such that a uniform biological response was achieved with a cell survival rate of 7% over the SOBP region. Biological experiments were then performed using the SOBP beam in a human salivary gland (HSG) cell line to measure the cell survival rates.

Results: The biological responses were uniform in the SOBP region, as expected by the MKM; however, the mean of the measured cell survival rates was (11.2 ± 0.6) % in the SOBP region, which was 60% higher than the designed rate. When investigating the biological parameters of the HSG cell line used in the experiments, we found that they were altered slightly from the MKM parameters used for carbon-ion radiotherapy. The new β parameter reproduced the measured survival rates within 6.5% in the SOBP region.

Conclusion: We produced biologically uniform SOBP using MKM for carbon-ion radiotherapy. The measured survival rates in the SOBP region were higher than expected, and the survival rates were reproduced by modifying the MKM parameter. This study was limited to one SOBP, and further investigations are required to prove that MKM is generally applicable to helium-ion radiotherapy.

Keywords: HSG cell line; Helium-ion therapy; Microdosimetric kinetic model; Ridge filter.