The evaluation of an on-site monitoring program for activity meter quality assurance with exemption-level sources

J Radiol Prot. 2023 May 12;43(2). doi: 10.1088/1361-6498/accc96.

Abstract

This paper discusses the feasibility of a monitoring program for the quality assurance status of activity meters. We sent a questionnaire to clinical nuclear medicine departments of medical institutions, requesting information on their activity meters and quality assurance practices. On-site visits were conducted with exemption-level standard sources (Co-57, Cs-137 and Ba-133) for dose calibrators in nuclear medicine departments including physical inspection, accuracy and reproducibility. A method offering a quick check on the detection efficiency of the space dimension inside the activity meters was also introduced. For dose calibrator quality assurance, the daily checks had the highest implementation. However, annual checks and upon acceptance/after a repair check were reduced to 50% and 44%, respectively. The accuracy results of dose calibrators showed that all models exceeded the ±10% criteria with Co-57 and Cs-137 sources. The reproducibility results showed that some models exceeded the ±5% criteria with Co-57 and Cs-137 sources. The appropriate application of exemption-level standard sources considering the uncertainty that affects the measurement is discussed.

Keywords: activity meter; exemption-level source; on-site visit; quality assurance; questionnaire.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cesium Radioisotopes*
  • Reproducibility of Results
  • Uncertainty

Substances

  • Cesium-137
  • Cesium Radioisotopes