Intravascular Sono-Ablation for In-Stent Restenosis Relief: Transducer Development and the In-Vitro Demonstration

IEEE Trans Biomed Eng. 2023 Jul;70(7):2172-2180. doi: 10.1109/TBME.2023.3238679. Epub 2023 Jun 19.

Abstract

Objective: This study aimed to propose a new clinical modality for the relief of in-stent restenosis (ISR) using focused ultrasound (FUS) ablation. In the first research stage, a miniaturized FUS device was developed for the sonification of the remaining plaque after stenting, known as one of the causes of ISR.

Methods: This study presents a miniaturized (<2.8 mm) intravascular FUS transducer for ISR treatment. The performance of the transducer was predicted through a structural-acoustic simulation, followed by fabrication of the prototype device. Using the prototype FUS transducer, we demonstrated tissue ablation with bio-tissues over metallic stents, mimicking in-stent tissue ablation. Next, we conducted a safety test by detecting the existence of thermal damage to the arterial tissue upon sonication with a controlled dose.

Results: The prototype device successfully delivered sufficient acoustic intensity (>30 W/cm2) to a bio tissue (chicken breast) through a metallic stent. The ablation volume was approximately 3.9 × 7.8 × 2.6 mm3. Furthermore, 1.5 min sonication was sufficient to obtain an ablating depth of approximately 1.0 mm, not thermally damaging the underlying artery vessel.

Conclusion: We demonstrated in-stent tissue sonoablation, suggesting it could be as a future ISR treatment modality.

Significance: Comprehensive test results provide a key understanding of FUS applications using metallic stents. Furthermore, the developed device can be used for sonoablation of the remaining plaque, providing a novel approach to the treatment of ISR.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Computer Simulation
  • Coronary Restenosis*
  • Humans
  • Stents
  • Treatment Outcome