Advanced Pipeline for Designing Multi-Locus TMS Coils With Current Density Constraints

IEEE Trans Biomed Eng. 2023 Jul;70(7):2025-2034. doi: 10.1109/TBME.2023.3234119. Epub 2023 Jun 19.

Abstract

Objective: This work aims for a method to design manufacturable windings for transcranial magnetic stimulation (TMS) coils with fine control over the induced electric field (E-field) distributions. Such TMS coils are required for multi-locus TMS (mTMS).

Methods: We introduce a new mTMS coil design workflow with increased flexibility in target E-field definition and faster computations compared to our previous method. We also incorporate custom current density and E-field fidelity constraints to ensure that the target E-fields are accurately reproduced with feasible winding densities in the resulting coil designs. We validated the method by designing, manufacturing, and characterizing a 2-coil mTMS transducer for focal rat brain stimulation.

Results: Applying the constraints reduced the computed maximum surface current densities from 15.4 and 6.6 kA/mm to the target value 4.7 kA/mm, yielding winding paths suitable for a 1.5-mm-diameter wire with 7-kA maximum currents while still replicating the target E-fields with the predefined 2.8% maximum error in the FOV. The optimization time was reduced by two thirds compared to our previous method.

Conclusion: The developed method allowed us to design a manufacturable, focal 2-coil mTMS transducer for rat TMS impossible to attain with our previous design workflow.

Significance: The presented workflow enables considerably faster design and manufacturing of previously unattainable mTMS transducers with increased control over the induced E-field distribution and winding density, opening new possibilities for brain research and clinical TMS.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Brain* / physiology
  • Head
  • Rats
  • Stereotaxic Techniques
  • Transcranial Magnetic Stimulation* / methods
  • Transducers