Flow-through charge detection mass spectrometry to categorize micron-sized particles

Rev Sci Instrum. 2023 Mar 1;94(3):034104. doi: 10.1063/5.0131941.

Abstract

This paper outlines the use of charge detection mass spectrometry to simultaneously measure the charge and mass of micron-sized particles. In a flow-through instrument, the detection of charge was achieved through charge induction onto cylindrical electrodes that connect to a differential amplifier. Mass was determined by particle acceleration under the influence of an electric field. Particles ranging from 30 to 400 fg (3 to 7 µm diameter) were tested. The detector design can measure particle mass within 10% accuracy for particles up to 620 fg with total charge ranging from 500e- to 56 ke-. This charge and mass range are expected to be relevant for dust on Mars.