Validation of web-based audiometry version of HEARZAP

PLoS One. 2023 Mar 28;18(3):e0283519. doi: 10.1371/journal.pone.0283519. eCollection 2023.

Abstract

Aim: The purpose of this study was to verify the accuracy of the web-based audiometer HEARZAP in determining hearing thresholds for both air and bone conduction.

Method: Using a cross-sectional validation design, the web-based audiometer was compared to a gold standard audiometer. Participants in the study totaled 50 (100 ears), of which 25 (50 ears) had normal hearing sensitivity and 25 (50 ears) had various types and degrees of hearing loss. All subjects underwent pure tone audiometry, including air and bone conduction thresholds, using the web-based and gold standard audiometers in a random order. A pause between the two tests was allowed if the patient felt comfortable. The testing for the web-based audiometer and gold standard audiometer was done by two different audiologists with similar qualifications in order to eliminate tester bias. Both the procedures were performed in a sound treated room.

Results: For air conduction thresholds and bone conduction thresholds, respectively, the mean discrepancies between the web-based audiometer and the gold standard audiometer were 1.22 dB HL (SD = 4.61) and 0.8 dB HL (SD = 4.1). The ICC for air conduction thresholds between the two techniques was 0.94 and for the bone conduction thresholds was 0.91. The Bland Altman plots likewise indicated excellent reliability between the two measurements, with the mean difference between the HEARZAP and the gold standard audiometry falling within the top and lower limits of agreement.

Conclusion: The web-based audiometry version of HEARZAP produced precise findings for hearing thresholds that were comparable to those obtained from an established gold standard audiometer. HEARZAP, has the potential to support multi-clinic functionality and enhance service access.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Audiometry*
  • Audiometry, Pure-Tone / methods
  • Auditory Threshold
  • Cross-Sectional Studies
  • Humans
  • Internet*
  • Reproducibility of Results

Grants and funding

The corresponding author received in-part funding support from Izameditechnologies PVT Ltd. for this study. The testing lab and equipment used for the study was funded through the corresponding author's grant 'The Wellcome Trust DBT India Alliance grant' (IA/CPHI/19/1/504614). The funders did not play any role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript.