A graphite transmission ionization chamber

Med Phys. 1987 Nov-Dec;14(6):1056-9. doi: 10.1118/1.596121.

Abstract

A pancake-type transmission chamber made of high-purity graphite and open to the atmosphere has been designed and constructed at the Secondary Standard Dosimetry Laboratory (SSDL-Rio de Janeiro). Tests performed on the chamber following the International Electrotechnical Commission recommendations indicate that its performance characteristics are comparable to those expected from a secondary standard ionization chamber.

MeSH terms

  • Cobalt Radioisotopes
  • Graphite
  • Humans
  • Radiation Dosage*
  • Radioisotope Teletherapy / instrumentation
  • Radiotherapy / instrumentation*
  • Radiotherapy Dosage*

Substances

  • Cobalt Radioisotopes
  • Graphite