Detection of Tip Fold-Over of the Slim Modiolar Electrode Using Intraoperative Mobile Cone-Beam Computed Tomography

Otolaryngol Head Neck Surg. 2023 Jul;169(1):105-111. doi: 10.1002/ohn.240. Epub 2023 Jan 25.

Abstract

Objective: This study aimed to evaluate the importance of mobile cone-beam computed tomography in detecting tip fold-over of a slim modiolar electrode within the cochlea during surgery.

Study design: Retrospective case review.

Setting: Tertiary medical center.

Methods: From January 2020 to June 2022, 33 ears of 30 patients with normal cochlear morphology underwent cochlear implantation with slim modiolar electrodes and intraoperative mobile cone-beam computed tomography imaging. Furthermore, we retrospectively reviewed the medical records and images.

Results: The tip fold-over of the electrodes was detected using mobile cone-beam computed tomography in 3 out of 33 ears (9.1%). We could not identify the tip fold-over by scouting plain X-ray images in 2 out of 3 cases before taking the cone-beam computed tomography images. Electrode removal and reinsertion were performed before wound closure and the successful reinsertion was confirmed by mobile cone-beam computed tomography. The folded electrode tips were located at 238.8°, 152°, and 185.8°.

Conclusion: Intraoperative mobile cone-beam computed tomography is useful in detecting the tip fold-over of the slim modiolar electrodes during surgery. Therefore, it was possible to reinsert the electrodes in all cases before closing the wound, eliminating the need for revision surgeries. Moreover, the analysis of mobile cone-beam computed tomography images may help to elucidate the mechanisms of electrode tip fold-over.

Keywords: cochlear implantation; cone-beam computed tomography; slim modiolar electrode; tip fold-over.