In-line warming reduces in-line pressure of subcutaneous infusion of concentrated immunoglobulins

Drug Deliv Transl Res. 2023 Sep;13(9):2367-2377. doi: 10.1007/s13346-023-01321-y. Epub 2023 Mar 15.

Abstract

Immunoglobulin replacement therapy is a life-saving treatment in patients with immunodeficiency and effective in the management of autoimmune disorders. Immunoglobulins are administered intravenously or subcutaneously, with the latter route reducing systemic reactions and providing an option for self-infusion, increasing patient convenience, while decreasing patient burden, healthcare utilization, and costs. A major limitation with subcutaneous administrations is the frequency of infusion due to limited volumes administrable into subcutaneous space, necessitating increased drug concentration, absorption, and dispersion. Increasing the concentration of immunoglobulins from 10 to 20% halves the required volume, but leads to higher dynamic viscosity, limiting infusion rate. Recombinant human hyaluronidase increases dispersion and absorption of immunoglobulins allowing administration of ≤ 600 mL per site, but does not change viscosity. Since the viscosity of fluids depends on temperature, we tested the feasibility of in-line warming of immunoglobulin formulations to physiological temperatures. In vitro analysis showed no negative impact of in-line warming to 38 °C on product quality. Subcutaneous infusion studies in pigs confirmed the feasibility of infusion rates of up to 7.5 mL/min with in-line warmed TAK-881, an immunoglobulin 20% facilitated with recombinant human hyaluronidase. In-line pressures were reduced compared with conventional immunoglobulin 20%, and local tolerance was not altered. Reduction of in-line pressures was more pronounced with thinner needle sets, indicating a potential benefit for patients. In summary, an in in-line warming device can circumvent the limitation of high viscosity, while product quality and local tolerance are maintained. The results of the presented studies warrant further testing in a phase 1 clinical study.

Keywords: CUVITRU; HYQVIA; Immunoglobulins; SCIG; SCIG 20%; Subcutaneous administration; TAK-881; fSCIG 20%; rHuPH20.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Humans
  • Hyaluronoglucosaminidase* / adverse effects
  • Immunoglobulins / adverse effects
  • Immunologic Deficiency Syndromes* / drug therapy
  • Infusions, Subcutaneous
  • Injections, Subcutaneous
  • Swine

Substances

  • Hyaluronoglucosaminidase
  • Immunoglobulins