Simulation studies of a full-ring, CZT SPECT system for whole-body imaging of 99m Tc and 177 Lu

Med Phys. 2023 Jun;50(6):3726-3737. doi: 10.1002/mp.16360. Epub 2023 Mar 22.

Abstract

Background: Single photon emission computed tomography (SPECT) is an imaging modality that has demonstrated its utility in a number of clinical indications. Despite this progress, a high sensitivity, high spatial resolution, multi-tracer SPECT with a large field of view suitable for whole-body imaging of a broad range of radiotracers for theranostics is not available.

Purpose: With the goal of filling this technological gap, we have designed a cadmium zinc telluride (CZT) full-ring SPECT scanner instrumented with a broad-energy tungsten collimator. The final purpose is to provide a multi-tracer solution for brain and whole-body imaging. Our static SPECT does not rely on the dual- and the triple-head rotational SPECT standard paradigm, enabling a larger effective area in each scan to increase the sensitivity. We provide a demonstration of the performance of our design using a realistic model of our detector with simulated body-sized phantoms filled with 99m Tc and 177 Lu.

Methods: We create a realistic model of our detector by using a combination of a Geant4 Application for Tomographic Emission (GATE) Monte Carlo simulation and a finite element model for the CZT response, accounting for low-energy tail effects in CZT that affects the sensitivity and the scatter correction. We implement a modified dual-energy-window scatter correction adapted for CZT. Other corrections for attenuation, detector and collimator response, and detector gaps and edges are also included. The images are reconstructed using the maximum-likelihood expectation-maximization. Detector and reconstruction performance are characterized with point sources, Derenzo phantoms, and a body-sized National Electrical Manufacturers Association (NEMA) Image Quality (IQ) phantom for both 99m Tc and 177 Lu.

Results: Our SPECT design can resolve 7.9 mm rods for 99m Tc (140 keV) and 9.5 mm for 177 Lu (208 keV) in a hot-rod Derenzo phantom with a 3-min exposure and reach an image contrast of 78% for 99m Tc and 57% for 177 Lu using the NEMA IQ phantom with a 6-min exposure. Our modified scatter correction shows an improved contrast-recovery ratio compared to a standard correction.

Conclusions: In this paper, we demonstrate the good performance of our design for whole-body imaging purposes. This adds to our previous demonstration of improved qualitative and quantitative 99m Tc imaging of brain perfusion and 123 I imaging of dopamine transport with respect to state-of-the-art NaI dual-head cameras. We show that our design provides similar IQ and contrast to the commercial full-ring SPECT VERITON for 99m Tc. Regarding 177 Lu imaging of the 208 keV emissions, our design provides similar contrast to that of other state-of-the-art SPECTs with a significant reduction in exposure. The high sensitivity and extended energy range up to 250 keV makes our SPECT design a promising alternative for clinical imaging and theranostics of emerging radionuclides.

Keywords: CZT; Monte Carlo simulation; SPECT; finite element method; full-ring.