Improved lateral penumbra for proton ocular treatments on a general-purpose spot scanning beamline

Phys Med. 2023 Mar:107:102551. doi: 10.1016/j.ejmp.2023.102551. Epub 2023 Mar 1.

Abstract

Purpose: An ocular applicator that fits a commercial proton snout with an upstream range shifter to allow for treatments with sharp lateral penumbra is described.

Materials and methods: The validation of the ocular applicator consisted of a comparison of range, depth doses (Bragg peaks and spread out Bragg peaks), point doses, and 2-D lateral profiles. Measurements were made for three field sizes, 1.5, 2, and 3 cm, resulting in 15 beams. Distal and lateral penumbras were simulated in the treatment planning system for seven range-modulation combinations for beams typical of ocular treatments and a field size of 1.5 cm, and penumbra values were compared to published literature.

Results: All the range errors were within 0.5 mm. The maximum averaged local dose differences for Bragg peaks and SOBPs were 2.6% and 1.1%, respectively. All the 30 measured point doses were within +/-3% of the calculated. The measured lateral profiles, analyzed through gamma index analysis and compared to the simulated, had pass rates greater than 96% for all the planes. The lateral penumbra increased linearly with depth, from 1.4 mm at 1 cm depth to 2.5 mm at 4 cm depth. The distal penumbra ranged from 3.6 to 4.4 mm and increased linearly with the range. The treatment time for a single 10 Gy (RBE) fractional dose ranged from 30 to 120 s, depending on the shape and size of the target.

Conclusions: The ocular applicator's modified design allows lateral penumbra similar to dedicated ocular beamlines while enabling planners to use modern treatment tools such as Monte Carlo and full CT-based planning with increased flexibility in beam placement.

Keywords: General-purpose beamline; Lateral penumbra; Ocular proton therapy; Pencil beam scanning.

MeSH terms

  • Monte Carlo Method
  • Phantoms, Imaging
  • Proton Therapy* / methods
  • Protons*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods
  • Synchrotrons

Substances

  • Protons