Flow Rate-Independent Multiscale Liquid Biopsy for Precision Oncology

ACS Sens. 2023 Mar 24;8(3):1200-1210. doi: 10.1021/acssensors.2c02577. Epub 2023 Feb 20.

Abstract

Immunoaffinity-based liquid biopsies of circulating tumor cells (CTCs) hold great promise for cancer management but typically suffer from low throughput, relative complexity, and postprocessing limitations. Here, we address these issues simultaneously by decoupling and independently optimizing the nano-, micro-, and macro-scales of an enrichment device that is simple to fabricate and operate. Unlike other affinity-based devices, our scalable mesh approach enables optimum capture conditions at any flow rate, as demonstrated with constant capture efficiencies, above 75% between 50 and 200 μL min-1. The device achieved 96% sensitivity and 100% specificity when used to detect CTCs in the blood of 79 cancer patients and 20 healthy controls. We demonstrate its postprocessing capacity with the identification of potential responders to immune checkpoint inhibition (ICI) therapy and the detection of HER2 positive breast cancer. The results compare well with other assays, including clinical standards. This suggests that our approach, which overcomes major limitations associated with affinity-based liquid biopsies, could help improve cancer management.

Keywords: circulating tumor cells; liquid biopsy; mass transport; microfluidics; multiscale; precision oncology.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Breast Neoplasms*
  • Cell Line, Tumor
  • Female
  • Humans
  • Liquid Biopsy
  • Neoplastic Cells, Circulating* / pathology
  • Precision Medicine