A detailed evaluation of the advantages and limitations of online RP-LC×HILIC compared to HILIC×RP-LC for phenolic analysis

J Chromatogr A. 2023 Mar 15:1692:463843. doi: 10.1016/j.chroma.2023.463843. Epub 2023 Feb 4.

Abstract

The combination of hydrophilic interaction chromatography (HILIC) and reversed-phase liquid chromatography (RP-LC) has proved effective in the LC × LC analysis of polyphenols due to the high degree of orthogonality associated with these separation modes for various classes of phenolic compounds. However, despite the growing number of such applications, HILIC is almost exclusively used as the first dimension (1D) separation mode, and RP-LC in the second dimension (2D). This is somewhat surprising in light of the potential advantages of swapping these separation modes. In this contribution, we present a detailed evaluation of the potential of online RP-LC × HILIC-MS for the analysis of phenolic compounds, comparing the performance of this system to the more established HILIC × RP-LC-MS configuration. Method development was performed using a predictive optimisation program, and fixed solvent modulation was employed to combat the solvent incompatibility between HILIC and RP-LC mobile phases. Red wine, rooibos tea, Protea and chestnut phenolic extracts containing a large diversity of phenolic compound classes were analysed by both HILIC × RP-LC- and RP-LC × HILIC-MS in order to compare the separation performance. Overall, the kinetic performance of HILIC × RP-LC was found to be clearly superior, with higher peak capacities and better resolution obtained for the majority of samples compared to RP-LC × HILIC analyses using similar column dimensions. Dilution of the 1D solvent combined with large volume injections proved insufficient to focus especially phenolic acids in the 2D HILIC separation, which resulted in severe 2D peak distortion for these compounds, and negatively impacted on method performance. On the other hand, a noteworthy improvement in the sensitivity of RP-LC × HILIC-MS analyses was observed due to higher ESI-MS response for the 2D HILIC mobile phase and greater sample loading capacity of the 1D RP-LC column, brought on by the high solubility of phenolic samples in aqueous solutions. As a result, a significantly higher number of compounds were detected in the RP-LC × HILIC-MS separations. These findings point to the potential advantage of RP-LC × HILIC as a complementary configuration to HILIC × RP-LC for phenolic analysis.

Keywords: Comprehensive two-dimensional liquid chromatography (LC × LC); High resolution mass spectrometry (HR-MS); Hydrophilic interaction chromatography (HILIC); Phenolics; Reversed phase liquid chromatography (RP-LC).

MeSH terms

  • Chromatography, Reverse-Phase* / methods
  • Hydrophobic and Hydrophilic Interactions
  • Phenols* / analysis
  • Solvents

Substances

  • Phenols
  • Solvents