Angiojet™ mechanical thrombectomy-induced hemolysis triggering blood-leak alarms

Hemodial Int. 2023 Apr;27(2):E19-E22. doi: 10.1111/hdi.13062. Epub 2023 Feb 3.

Abstract

Blood leak alarms are increasingly rare and may be triggered by rupture of the dialyzer membrane, or intravascular hemolysis. We report two patients who developed hemolysis following Angiojet™ thrombolysis and thrombectomy of thrombosed arteriovenous fistulas, triggering blood leak alarms on hemodialysis, the occurrence of which has not been reported before. AngioJet™-induced hemolysis should be considered in the differential diagnosis for blood leak alarms occurring soon after an intervention on an arterio-venous (AV) fistula.

Keywords: Angiojet™ thrombolysis and thrombectomy; blood leak alarms; hemodialysis; hemolysis.

Publication types

  • Case Reports

MeSH terms

  • Arteriovenous Shunt, Surgical*
  • Hemolysis
  • Humans
  • Renal Dialysis* / adverse effects
  • Thrombectomy / adverse effects
  • Treatment Outcome
  • Vascular Patency