Comparison of perioperative outcomes between bipolar sealing, ultrasonic shears and a hybrid device during laparoscopic gastrectomy for early gastric cancer: a prospective, multicenter, randomized study

Gastric Cancer. 2023 May;26(3):438-450. doi: 10.1007/s10120-023-01365-6. Epub 2023 Feb 3.

Abstract

Background: Although EBDs are essential for minimally invasive surgery, well-established prospective randomized studies comparing EBDs are scarce. This study aimed to compare the intraoperative inflammatory response and short-term surgical outcomes among different energy-based devices (EBDs) in laparoscopic distal gastrectomy (LDG).

Methods: Patients with clinical stage I gastric cancer scheduled for LDG at two different medical centers were prospectively randomized into three groups: ultrasonic shears (US), advanced bipolar (BP) and ultrasonic-bipolar hybrid (HB). The C-reactive protein (CRP) level, operation time, intraoperative blood loss (IBL), laboratory tests, cytokines (interleukin (IL)-6 and IL-10), hospital stay, and complication rate were analyzed. A novel semiquantitative measurement method using indocyanine green (ICG) and a near-infrared camera measured the amount of lymphatic leakage.

Results: The primary endpoint, the CRP level, was significantly lower in the BP (n = 60) group than in the US (n = 57) or HB (n = 57) group [9.03 ± 5.55 vs. 11.12 ± 5.02 vs. 12.67 ± 6.14, p = 0.001, on postoperative day (POD) 2 and 7.48 vs. 9.62 vs. 9.48, p = 0.026, on POD 4]. IBL was significantly lower in BP than in US or HB (26.3 ± 25.3 vs. 43.7 ± 42.0 vs. 34.9 ± 37.0, p = 0.032). Jackson-Pratt drainage triglycerides were significantly lower in BP than in US (53.6 ± 33.7 vs. 84.2 ± 59.0, p = 0.11; HB: 71.3 ± 51.4). ICG fluorescence intensity, operation time, laboratory results, cytokines, hospital stay, and complication rate were not significantly different among the 3 groups.

Conclusion: BP showed a lower postoperative CRP level and less IBL than US and HB, suggesting less collateral thermal damage and better sealing function. Surgeons may consider this when selecting EBDs for laparoscopic surgery.

Keywords: Bipolar; Early gastric cancer; Energy devices; Laparoscopy gastrectomy; Ultrasonic.

Publication types

  • Randomized Controlled Trial
  • Multicenter Study

MeSH terms

  • Gastrectomy / methods
  • Humans
  • Indocyanine Green
  • Interleukin-10
  • Laparoscopy* / methods
  • Prospective Studies
  • Retrospective Studies
  • Stomach Neoplasms* / surgery
  • Treatment Outcome
  • Ultrasonics