Corneal thickness measurements with the RTVue, Casia-2, and Pentacam devices in patients with mild-to-moderate keratoconus: a comparative study

BMC Ophthalmol. 2023 Jan 26;23(1):36. doi: 10.1186/s12886-023-02767-x.

Abstract

Background: To compare the characteristics of corneal thickness measurements among the RTVue, Casia-2, and Pentacam in patients with mild-to-moderate keratoconus.

Methods: We recruited 46 eyes of 46 patients diagnosed with mild-to-moderate keratoconus at our hospital between January and March 2022. The central corneal thickness (CCT) and thinnest corneal thickness (TCT) were measured using two optical coherence tomography (OCT) instruments (RTVue and Casia-2) and the more conventional Pentacam. Differences and correlations between the CCTs and TCTs, based on the device and influencing factors, were explored.

Results: The CCTs were highly consistent among the groups (p = 0.434) and correlated with one another (p < 0.001). The TCTs measured by OCTs were thinner than those measured by the Pentacam (p < 0.001); however, all three devices were highly correlated (p < 0.001). The thinnest point location measurements with RTVue and Casia-2 differed significantly from the measurements with the Pentacam. Bland-Altman plots demonstrated a significant agreement between Pentacam and OCTs in TCT measurement (p < 0.001); the 95% limits of agreement were - 3.1 μm to + 33.1 μm for Pentacam and RTVue and - 8.6 μm to + 36.5 μm for Pentacam and Casia-2. RTVue and Casia-2 showed no difference in corneal thickness (p = 0.633) and thinnest point location measurement (p > 0.05). Multivariate analysis identified that the TCT measurement difference between the RTVue and Pentacam was related to the difference between the CCT and TCT (b = 0.490, 95% confidence interval [CI]: 0.033 to 0.948, p = 0.036), whereas the difference between the Casia-2 and Pentacam was related to the anterior radius for curvature (A) grade (b = 3.9, 95% CI: 1.753 to 6.074, p = 0.001), corneal pachymetry at the thinnest (C) grade (b = - 7.875, 95% CI: - 11.404 to - 4.346, p < 0.001), and the difference between the CCT and TCT (b = 0.425, 95% CI: 0.1 to 0.751, p = 0.012).

Conclusions: CCTs in patients with mild-to-moderate keratoconus were similar among all three devices, but the TCTs and the thinnest point locations were not. Furthermore, the TCT measurement differences between the OCT devices and the Pentacam were more pronounced in keratoconus cases with a steeper anterior surface, thicker TCTs, and a larger difference between the CCT and TCT.

Trial registration: Number: 2021118-1. Retrospectively registered: September 01, 2021.

Keywords: Agreement; Corneal thickness; Keratoconus; Optical coherence tomography.

MeSH terms

  • Cornea
  • Corneal Pachymetry / methods
  • Corneal Topography / methods
  • Humans
  • Keratoconus* / diagnosis
  • Prospective Studies
  • Reproducibility of Results
  • Tomography, Optical Coherence / methods