Diagnostic compatibility of various fixed orthodontic retainers for head/neck MRI and dental MRI

Clin Oral Investig. 2023 May;27(5):2375-2384. doi: 10.1007/s00784-023-04861-2. Epub 2023 Jan 14.

Abstract

Objectives: To evaluate the diagnostic MRI compatibility of different fixed orthodontic retainers using a high-resolution 3D-sequence optimized for artifact reduction.

Materials and methods: Maxillary and mandibular retainers made of five different materials were scanned in vitro and in vivo at 3 T MRI using an MSVAT-SPACE sequence. In vitro, artifact volumes were determined for all maxillary and mandibular retainers (AVmax; AVmand). In vivo, two independent observers quantified the extent of artifacts based on the visibility of 124 dental and non-dental landmarks using a five-point rating scale (1 = excellent, 2 = good, 3 = acceptable, 4 = poor, 5 = not visible).

Results: Rectangular-steel retainers caused the largest artifacts (AVmax/AVmand: 18,060/15,879 mm3) and considerable diagnostic impairment in vivo (mean landmark visibility score ± SD inside/outside the retainer areas: 4.8 ± 0.8/2.9 ± 1.6). Smaller, but diagnostically relevant artifacts were observed for twistflex steel retainers (437/6317 mm3, 3.1 ± 1.7/1.3 ± 0.7). All retainers made of precious-alloy materials produced only very small artifact volumes (titanium grade 1: 70/46 mm3, titanium grade 5: 47/35 mm3, gold: 23/21 mm3) without any impact on image quality in vivo (each retainer: visibility scores of 1.0 ± 0.0 for all landmarks inside and outside the retainer areas).

Conclusions: In contrast to steel retainers, titanium and gold retainers are fully compatible for both head/neck and dental MRI when using MSVAT-SPACE.

Clinical relevance: This study demonstrates that titanium and gold retainers do not impair the diagnostic quality of head/neck and dental MRI when applying an appropriate artifact-reduction technique. Steel retainers, however, are not suitable for dental MRI and can severely impair image quality in head/neck MRI of the oral cavity.

Keywords: Artifacts; Dentistry; Diagnostic imaging; Magnetic resonance imaging; Orthodontic retainers; Orthodontics.

MeSH terms

  • Gold
  • Magnetic Resonance Imaging / methods
  • Mouth
  • Orthodontic Retainers*
  • Stainless Steel
  • Titanium*

Substances

  • Titanium
  • Stainless Steel