Reliable Fast (20 Hz) Acquisition Rate by a TD fNIRS Device: Brain Resting-State Oscillation Studies

Sensors (Basel). 2022 Dec 24;23(1):196. doi: 10.3390/s23010196.

Abstract

A high power setup for multichannel time-domain (TD) functional near infrared spectroscopy (fNIRS) measurements with high efficiency detection system was developed. It was fully characterized based on international performance assessment protocols for diffuse optics instruments, showing an improvement of the signal-to-noise ratio (SNR) with respect to previous analogue devices, and allowing acquisition of signals with sampling rate up to 20 Hz and source-detector distance up to 5 cm. A resting-state measurement on the motor cortex of a healthy volunteer was performed with an acquisition rate of 20 Hz at a 4 cm source-detector distance. The power spectrum for the cortical oxy- and deoxyhemoglobin is also provided.

Keywords: brain; diffuse optics; functional near infrared spectroscopy; hemodynamics; resting-state brain oscillation; time domain.

MeSH terms

  • Brain / diagnostic imaging
  • Brain Mapping / methods
  • Humans
  • Motor Cortex* / diagnostic imaging
  • Signal-To-Noise Ratio
  • Spectroscopy, Near-Infrared* / methods