Accuracy of fully guided dental implant placement: A prospective clinical in-vivo investigation using intraoral scan data

Int J Comput Dent. 2023 May 26;26(2):137-148. doi: 10.3290/j.ijcd.b3762753.

Abstract

Aim: To compare the planned implant position (PIP) with the transferred implant position (TIP) after fully guided implant placements in single-tooth gaps.

Materials and methods: Dental implant placements were planned using two different implant systems (Camlog Screw-Line [C-SL] and Straumann Bone Level Tapered [S-BLT]), and two different planning software programs (SMOP and coDiagnostiX). All implants were placed according to fully guided protocols, and intraoral scans were performed intraoperatively. For the comparison of PIP and TIP, scan data were imported to Geomagic Control X (GCX) software and accuracies were evaluated. Deviations were reported in a coordinate system (x- [mesiodistal], y- [vestibulo-oral], and z- [vertical] axis) at entry points and apices. Total deviations, including angular deviations, were calculated with GCX. For statistical analysis, the level of significance was set to P < 0.05.

Results: Twenty-six patients received 26 implants. Mean 3D deviation at the implant's entry point was 0.61 mm ± 0.28 for C-SL and 0.63 mm ± 0.24 for S-BLT. For the implant's apex, mean 3D deviation of 0.96 mm ± 0.41 was documented for C-SL and 1.04 mm ± 0.34 for S-BLT. Mean angular deviation was 2.58 degrees ± 1.40 for C-SL and 2.89 degrees ± 1.12 for S-BLT. Statistical analysis revealed no significant differences between implant systems, but showed significant deviations regarding the z-axis, both at entry point and apex (P < 0.05).

Conclusions: Fully guided implant placements in single-tooth gaps provide accurate results. Due to significant vertical deviations, reevaluation of both drilling and insertion depths prior to implant installation should be considered. Maintenance of 1.5- to 2-mm safety distances to critical structures was confirmed.

Keywords: CAD/CAM; backward planning; computer-assisted; computer-assisted implant surgery; dental implants; static navigation.

MeSH terms

  • Computer-Aided Design
  • Cone-Beam Computed Tomography
  • Dental Implantation, Endosseous / methods
  • Dental Implants*
  • Humans
  • Imaging, Three-Dimensional
  • Prospective Studies
  • Surgery, Computer-Assisted*

Substances

  • Dental Implants