Abstract

Purpose: Analysis of cochlear structures and postoperative temporal bone (TB) imaging are gaining importance in the evaluation of cochlear implantation (CI°). Our aims were to explore the microarchitecture of human cochlea using micro-computed tomography (μCT), analyze electrode's placement inside cochlea after CI°, and compare pre-/post-implantation μCT scans with cone-beam CT (CBCT) scans of same TBs.

Methods: Cadaveric TBs were scanned using μCT and CBCT then underwent CI° using straight electrodes. Thereafter, they underwent again μCT and CBCT-imaging.

Results: Ten TBs were studied. μCT allowed visualization of scala tympani, scala vestibuli, basilar membrane, osseous spiral lamina, crista fenestrae, and spiral ligament. CBCT showed same structures except spiral ligament and crista fenestrae. After CI°, μCT and CBCT displayed the scalar location and course of electrode array within the cochlea. There were 7 cases of atraumatic electrode insertion and 3 cases of insertion trauma: basilar membrane elevation, electrode foldover with limited migration into scala vestibuli, and electrode kinking with limited migration into scala vestibuli. Insertion trauma was not correlated with cochlea's size or crista's maximal height but with round window membrane diameter. Resolution of μCT was higher than CBCT but electrode artifacts were similar.

Conclusions: μCT was accurate in visualizing cochlear structures, and course and scalar position of electrode array inside cochlea with any possible trauma to cochlea or array. CBCT offers a good alternative to μCT in clinical practice for cochlear imaging and evaluation of CI°, with lower radiation and higher resolution than multi-slice CT. Difficulties related to non-traumatic CI° are multifactorial.

Keywords: Cochlea; Cochlear implantation; Cochlear trauma; Cone-beam CT; Micro-CT; Temporal bone histology.

MeSH terms

  • Cochlea / diagnostic imaging
  • Cochlea / surgery
  • Cochlear Implantation* / methods
  • Cochlear Implants*
  • Cone-Beam Computed Tomography
  • Humans
  • Temporal Bone / diagnostic imaging
  • Temporal Bone / surgery
  • X-Ray Microtomography