Precise quantum measurement of vacuum with cold atoms

Rev Sci Instrum. 2022 Dec 1;93(12):121101. doi: 10.1063/5.0120500.

Abstract

We describe the cold-atom vacuum standards (CAVS) under development at the National Institute of Standards and Technology (NIST). The CAVS measures pressure in the ultra-high and extreme-high vacuum regimes by measuring the loss rate of sub-millikelvin sensor atoms from a magnetic trap. Ab initio quantum scattering calculations of cross sections and rate coefficients relate the density of background gas molecules or atoms to the loss rate of ultra-cold sensor atoms. The resulting measurement of pressure through the ideal gas law is traceable to the second and the kelvin, making it a primary realization of the pascal. At NIST, two versions of the CAVS have been constructed: a laboratory standard used to achieve the lowest possible uncertainties and pressures, and a portable version that is a potential replacement for the Bayard-Alpert ionization gauge. Both types of CAVSs are connected to a combined extreme-high vacuum flowmeter and dynamic expansion system to enable sensing of a known pressure of gas. In the near future, we anticipate being able to compare the laboratory scale CAVS, the portable CAVS, and the flowmeter/dynamic expansion system to validate the operation of the CAVS as both a standard and vacuum gauge.