A Comparative Study on a Novel Quality Assessment Protocol Based on Image Analysis Methods for Color Doppler Ultrasound Diagnostic Systems

Sensors (Basel). 2022 Dec 15;22(24):9868. doi: 10.3390/s22249868.

Abstract

Color Doppler (CD) imaging is widely used in diagnostics since it allows real-time detection and display of blood flow superimposed on the B-mode image. Nevertheless, to date, a shared worldwide standard on Doppler equipment testing is still lacking. In this context, the study herein proposed would give a contribution focusing on the combination of five test parameters to be included in a novel Quality Assessment (QA) protocol for CD systems testing. A first approach involving the use of the Kiviat diagram was investigated, assuming the diagram area, normalized with respect to one of the gold standards, as an index of the overall Doppler system performance. The QA parameters were obtained from the post-processing of CD data through the implementation of custom-written image analysis methods and procedures, here applied to three brand-new high-technology-level ultrasound systems. Experimental data were collected through phased and convex array probes, in two configuration settings, by means of a Doppler flow phantom set at different flow rate regimes. The outcomes confirmed that the Kiviat diagram might be a promising tool applied to quality controls of Doppler equipment, although further investigations should be performed to assess the sensitivity and specificity of the proposed approach.

Keywords: Kiviat diagram; color doppler; flow phantom; image analysis methods; quality assessment; test parameters; ultrasound diagnostic systems.

MeSH terms

  • Hemodynamics*
  • Phantoms, Imaging
  • Sensitivity and Specificity
  • Ultrasonography / methods
  • Ultrasonography, Doppler, Color*

Grants and funding

This research received no external funding.