Background: The aim of the study was to investigate the effect of three aligner cleaners on the composition and mechanical properties of two types of orthodontic aligners.
Materials and methods: The cleaners tested were two alkaline peroxide solutions (Retainer Brite-RB; Retainer Cleaner-RC) and one peroxide-free (Steraligner-ST) and the aligners Clear Aligner (C, polyester) and Invisalign (I, polyester-urethane). The aligners were immersed in the cleaner solutions as instructed every day (15 min for RB, RC; 5 min for ST) for a two-week period. The acidity of the solutions was tested with a pH meter. The changes in the chemical composition of the aligners were studied by attenuated total-reflection Fourier transform infrared spectrometry (ATR-FTIR), while Instrumented Indentation Testing (IIT) was used for assessment of changes in Martens Hardness (HM), modulus (EIT), elastic index (nIT) and relaxation (RIT).
Results: RB and RC were weakly acidic (pH = 6.3), whereas ST was mildly acidic (pH = 4.8). The ATR-FTIR analysis demonstrated evidence of acidic hydrolysis of C in ST and I in RB. The IIT-derived properties of I were not affected by the cleaners. However, for C a significant change was found in HM (all cleaners), nIT (all cleaners) and RIT (RB, ST). Although the chemical changes support a hydrolytic material deterioration, the results of mechanical properties may interfere with the material residual stresses during fabrication.
Conclusions: Caution should be exerted in the selection of aligner cleaners. The mild acidic cleanser was more aggressive to the polyester, whereas an alkaline peroxide to the polyester-urethane aligner.
Keywords: Aligners; Chemical alterations; Cleansers; Mechanical properties.