Deformed dose restoration to account for tumor deformation and position changes for adaptive proton therapy

Med Phys. 2023 Feb;50(2):675-687. doi: 10.1002/mp.16149. Epub 2022 Dec 29.

Abstract

Background: Online adaptation during intensity-modulated proton therapy (IMPT) can minimize the effect of inter-fractional anatomical changes, but remains challenging because of the complex workflow. One approach for fast and automated online IMPT adaptation is dose restoration, which restores the initial dose distribution on the updated anatomy. However, this method may fail in cases where tumor deformation or position changes occur.

Purpose: To develop a fast and robust IMPT online adaptation method named "deformed dose restoration (DDR)" that can adjust for inter-fractional tumor deformation and position changes.

Methods: The DDR method comprises two steps: (1) calculation of the deformed dose distribution, and (2) restoration of the deformed dose distribution. First, the deformable image registration (DIR) between the initial clinical target volume (CTV) and the new CTV were performed to calculate the vector field. To ensure robustness for setup and range uncertainty and the ability to restore the deformed dose distribution, an expanded CTV-based registration to maintain the dose gradient outside the CTV was developed. The deformed dose distribution was obtained by applying the vector field to the initial dose distribution. Then, the voxel-by-voxel dose difference optimization was performed to calculate beam parameters that restore the deformed dose distribution on the updated anatomy. The optimization function was the sum of total dose differences and dose differences of each field to restore the initial dose overlap of each field. This method only requires target contouring, which eliminates the need for organs at risk (OARs) contouring. Six clinical cases wherein the tumor deformation and/or position changed on repeated CTs were selected. DDR feasibility was evaluated by comparing the results with those from three other strategies, namely, not adapted (continuing the initial plan), adapted by previous dose restoration, and fully optimized.

Results: In all cases, continuing the initial plan was largely distorted on the repeated CTs and the dose-volume histogram (DVH) metrics for the target were reduced due to the tumor deformation or position changes. On the other hand, DDR improved DVH metrics for the target to the same level as the initial dose distribution. Dose increase was seen for some OARs because tumor growth had reduced the relative distance between CTVs and OARs. Robustness evaluation for setup and range uncertainty (3 mm/3.5%) showed that deviation in DVH-bandwidth for CTV D95% from the initial plan was 0.4% ± 0.5% (Mean ± S.D.) for DDR. The calculation time was 8.1 ± 6.4 min.

Conclusions: An online adaptation algorithm was developed that improved the treatment quality for inter-fractional anatomical changes and retained robustness for intra-fractional setup and range uncertainty. The main advantage of this method is that it only requires target contouring alone and saves the time for OARs contouring. The fast and robust adaptation method for tumor deformation and position changes described here can reduce the need for offline adaptation and improve treatment efficiency.

Keywords: adaptive; deformed dose restoration; proton therapy.

MeSH terms

  • Humans
  • Neoplasms* / diagnostic imaging
  • Neoplasms* / radiotherapy
  • Organs at Risk
  • Proton Therapy* / methods
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods
  • Radiotherapy, Intensity-Modulated* / methods

Grants and funding