Three-year follow-up of aortic arch endovascular stent grafting with the Nexus device: results from a prospective multicentre study

Eur J Cardiothorac Surg. 2022 Dec 2;63(1):ezac561. doi: 10.1093/ejcts/ezac561.

Abstract

Objectives: Endovascular aortic arch stent grafting with branched devices has shown initial promising results. The aim of this prospective, multicentre study was to evaluate 3-year outcomes of aortic arch stent grafting with NEXUS® Aortic Arch Stent Graft System (Nexus), a single-branch, bi-modular, off-the-shelf aortic arch stent graft system in high-risk patients.

Methods: Patients treated with Nexus, either under the feasibility clinical study or as compassionate use procedures in 5 centres, were included in this study. The primary end point was overall survival. The secondary end points included the incidence of procedure-related unplanned intervention, stroke, paraplegia and endoleak. Clinical and radiologic follow-up was performed at each study site at 30 days, 6 months and on a yearly basis thereafter up to 3 years postoperatively.

Results: We analysed data from a total of 28 patients. The overall median follow-up was 1132 (interquartile range: 809-1537). There were no device or procedure-related deaths between 1 and 3 years. Overall survival at 1 and 3 years was 89% and 71%, respectively. The cumulative incidence of unplanned reintervention at 1 and 3 years was 11% and 29%, respectively. There were no reports of stroke, paraplegia, aneurysm rupture, myocardial infarction or new aortic valve insufficiency. In this study's 1-3 year follow-up period, 1 type Ib (4%), 1 type II (4%) and 2 type III (8%; between Nexus' distal end and Thoracic endovascular aortic repair (TEVAR) extensions) endoleak were detected.

Conclusions: Endovascular aortic arch exclusion with the single-branch, off-the-shelf Nexus system provides promising clinical and radiologic results at 3-year follow-up in a high-risk patient cohort.

Keywords: Aortic arch; Endovascular surgery; TEVAR.

Publication types

  • Multicenter Study

MeSH terms

  • Aortic Aneurysm, Thoracic*
  • Aortic Dissection*
  • Blood Vessel Prosthesis / adverse effects
  • Blood Vessel Prosthesis Implantation*
  • Endoleak / epidemiology
  • Endoleak / etiology
  • Endovascular Aneurysm Repair
  • Endovascular Procedures*
  • Follow-Up Studies
  • Humans
  • Postoperative Complications
  • Prospective Studies
  • Retrospective Studies
  • Stents / adverse effects
  • Stroke* / etiology
  • Treatment Outcome

Grants and funding