Mechanical Evaluation of Salvage Fixation Techniques for a Lateralized Fibular Tunnel in Posterolateral Corner Reconstruction of the Knee

Orthop J Sports Med. 2022 Nov 30;10(11):23259671221131303. doi: 10.1177/23259671221131303. eCollection 2022 Nov.

Abstract

Background: Posterolateral corner (PLC) reconstruction of the knee involves precise drilling of a single tunnel from anterolateral to posteromedial in the fibular head (FH) to ensure adequate graft fixation. Misplacement of the tunnel in a too lateral or too superior trajectory can lead to cortical breach and inadequate graft fixation.

Purpose: To (1) determine the mechanical consequence of a lateralized FH tunnel in PLC reconstruction and (2) compare the mechanical strength of 3 salvage fixation techniques for a lateralized FH tunnel.

Study design: Controlled laboratory study.

Methods: Sawbones models of a uniform density were utilized. FH tunnels (7-mm diameter) were drilled from anterolateral to posteromedial in positive controls (lateral cortex thickness, 7.6 ± 0.7 mm) to represent an improperly placed FH tunnel at risk of lateral cortical breach. For negative controls and salvage experimental groups, FH tunnels were drilled from anterolateral to posterolateral (lateral cortex thickness, 2.7 ± 0.9 mm). Three salvage fixation techniques were compared: suture anchor fixation, tunnel redrilling, and nitinol staple fixation. Samples (n = 5 per group) underwent uniaxial tension testing, and the ultimate tensile strength (UTS) and mode of failure were recorded. Data were analyzed using the 1-sample t test and nonparametric 1-sample Wilcoxon signed-rank test.

Results: The negative control group had a 4-fold lower mean UTS relative to the positive control group (1.49 ± 0.17 vs 6.25 ± 1.98 MPa; P < .01) and exhibited failure through the lateral cortex and tunnel. Nitinol staple fixation improved the mean UTS by >16 times compared with the negative control group (24.06 ± 6.49 vs 1.49 ± 0.17 MPa; P < .01). Suture anchors and tunnel redrilling exhibited similar UTS and mode of failure to those of negative controls.

Conclusion: Reinforcement of a thinned lateral FH cortex with a single nitinol staple improved graft fixation strength in a sawbones model.

Clinical relevance: A lateralized FH tunnel can be a common intraoperative pitfall during PLC reconstruction. Salvage of a thinned lateral FH cortex with a single nitinol staple may reduce the risk of cortical breach and graft failure.

Keywords: fibular head; ligament reconstruction; posterolateral corner; staple.