Initial experience on abdominal photon-counting computed tomography in clinical routine: general image quality and dose exposure

Eur Radiol. 2023 Apr;33(4):2461-2468. doi: 10.1007/s00330-022-09278-1. Epub 2022 Dec 8.

Abstract

Objectives: Photon-counting computed tomography has lately found its way into clinical routine. The new technique could offer substantial improvements regarding general image quality, image noise, and radiation dose reduction. This study evaluated the first abdominal examinations in clinical routine and compared the results to conventional computed tomography.

Methods: In this single-center retrospective study, 66 patients underwent photon-counting and conventional abdominal CT. Four radiologists assessed general image quality, image noise, and image artifacts. Signal-to-noise ratio and dose properties of both techniques within the clinical application were compared. An ex vivo phantom study revealed the radiobiological impact by means of DNA double-strand break foci in peripheral blood cells by enumerating γ-H2AX+53BP1 foci.

Results: General image quality in accordance with the Likert scale was found superior for photon-counting CT (4.74 ± 0.46 vs. 4.25 ± 0.54; p < 0.001). Signal-to-noise ratio (p < 0.001) and also dose exposure were higher for photon-counting CT (DLP: 419.2 ± 162.2 vs. 372.3 ± 236.6 mGy*cm; p = 0.0435). CT exposure resulted in significantly increased DNA damage in comparison to sham control (p < 0.001). Investigation of the average foci per cell and radiation-induced foci numbers revealed significantly elevated numbers (p = 0.004 and p < 0.0001, respectively) after photon-counting CT.

Conclusion: Photon-counting CT in abdominal examinations showed superior results regarding general image quality and signal-to-noise ratio in clinical routine. However, this seems to be traded for a significantly higher dose exposure and corresponding double-strand break frequency. Optimization of standard protocols in further clinical applications is required to find a compromise regarding picture quality and dose exposure.

Key points: • Photon-counting computed tomography promises to enhance the diagnostic potential of medical imaging in clinical routine. • Retrospective single-center study showed superior general image quality accompanied by higher dose exposure in initial abdominal PCCT protocols compared to state-of-the-art conventional CT. • A simultaneous ex vivo phantom study revealed correspondingly increased frequencies of DNA double-strand breaks after PCCT.

Keywords: DNA damage; Dose reduction; Signal-to-noise ratio; Tomography.

MeSH terms

  • DNA*
  • Humans
  • Phantoms, Imaging
  • Radiation Dosage
  • Retrospective Studies
  • Signal-To-Noise Ratio
  • Tomography, X-Ray Computed* / methods

Substances

  • DNA