Pedestal fluctuation measurements with charge exchange imaging at the DIII-D tokamak

Rev Sci Instrum. 2022 Nov 1;93(11):113503. doi: 10.1063/5.0101844.

Abstract

A new high radial resolution 2D multichannel Charge eXchange Imaging (CXI) diagnostic is under development for deployment at DIII-D. The diagnostic system will measure low-to-intermediate radial wavenumber carbon density fluctuations by observing the n = 8 - 7 (λ = 529.06 nm) C-VI emission line, resulting from charge exchange collisions between heating neutral beam atoms and the intrinsic carbon ion density. The new CXI diagnostic will provide measurements with ΔR ∼ 0.4 cm to access higher kr instabilities (kr < 8 cm-1) predicted to arise in the steep-gradient region of the H-mode pedestal. The CXI system will feature 60 fiber bundles in a 12 × 5 arrangement, with each bundle consisting of four 1 mm fibers. A custom optical system has been designed to filter and image incoming signals onto an 8 × 8 avalanche photodiode array. Additionally, a novel electronics suite has been designed and commissioned to amplify and digitize the relatively low-intensity carbon signal at a 2 MHz bandwidth. Forward modeling results of the active C-VI emission suggest sufficient signal to noise ratios to resolve turbulent fluctuations. Prototype measurements demonstrate the ability to perform high frequency pedestal measurements.