No difference in implant survival between 28-mm M2a RingLoc metal-on-metal and metal-on-polyethylene total hip arthroplasty: results from the Finnish Arthroplasty Register

Acta Orthop. 2022 Nov 24:93:854-858. doi: 10.2340/17453674.2022.5252.

Abstract

Background and purpose: Long-term outcome of small head (28 mm) metal-on-metal (MoM) total hip arthroplasty (THA) is available mainly for Metasul devices (Sulzer Medica, Winterthur, Switzerland). Biomet MoM THA was frequently used in Finland. Therefore, we assessed long-term survivorship of the M2a 28-mm RingLoc MoM THA (Biomet, Warsaw, IN, USA) and compared it with the metal-on-polyethylene (MoP) RingLoc THA from the same manufacturer.

Patients and methods: We conducted a register study based on THAs from the Finnish Arthroplasty Register performed between January 1, 2000 and December 31, 2007. 290 28-mm head M2a MoM THAs and 1,647 28-mm head MoP THAs (reference group) were included. The endpoint was revision for any reason, or revision for aseptic loosening, osteolysis, liner wear, or metallosis as one group. Kaplan-Meier survival estimates were calculated, and revision risks were assessed using a Cox multiple regression model, both with 95% confidence intervals (CI).

Results: No difference was found in the 15-year Kaplan-Meier survivorship between the 28-mm head M2a RingLoc MoM THA and the reference group for any reason for revision (87.7% [82.9-92.1] and 83.3% [81.0-85.3], respectively). The adjusted hazard ratio (HR) for any reason for revision for the MoM THA group compared with the reference group was at least equal or better (0.70 [0.48-1.02]). Both groups presented similar survival for revision for aseptic loosening of the cup, osteolysis, liner wear, or metallosis, at 96.2% (92.7-98.0) and 95.4% (93.9-96.5), respectively.

Interpretation: In the long-term survival there was no difference between the M2a 28-mm RingLoc MoM THA and 28-mm MoP THA. Further follow-up regimens for M2a 28-mm RingLoc THA patients may be unnecessary, but long-term metal ion and radiological data is needed before any formal suggestions.

MeSH terms

  • Arthroplasty, Replacement, Hip* / adverse effects
  • Cimetidine
  • Finland / epidemiology
  • Humans
  • Metal-on-Metal Joint Prostheses* / adverse effects
  • Metals
  • Osteolysis*
  • Polyethylene

Substances

  • Polyethylene
  • Cimetidine
  • Metals