First-in-Human Experience of Using a Universal Steerable Sheath in Implanting Left Atrial Appendage Closure Devices

JACC Asia. 2022 Nov 15;2(6):780-783. doi: 10.1016/j.jacasi.2022.08.003. eCollection 2022 Nov.

Abstract

Left atrial appendage (LAA) closure (LAAC) device implantation may be challenging in cases with difficult LAA anatomy. A deflectable sheath to approach multiple LAA positions may be helpful. We used a deflectable sheath to implant LAAC devices in 20 patients and included 60 cases using the standard sheaths for comparison. The procedures were successful in all patients without peri-procedural complications. After a median follow-up of 1.52 (IQR: 0.76-2.33) years, none of the patients in the deflectable sheath group had peri-device leak ≥3 mm and experienced thromboembolic stroke. In the standard sheath group, after follow-up of 2.03 (IQR: 0.87-3.19) years, 2 had peri-device leak ≥3 mm and 1 experienced thromboembolic stroke. We first proved the idea of using a universal steerable sheath for LAAC device implantation in difficult LAA anatomy, which also allows rapid switching of different LAAC devices.

Keywords: AF, atrial fibrillation; LAA, left atrial appendage; LAAC, left atrial appendage closure; deflectable sheath; left atrial appendage closure device; steerable sheath.