Diagnostic Accuracy of Frequency-Doubling Technology and the Moorfields Motion Displacement Test for Glaucoma

Ophthalmol Glaucoma. 2023 May-Jun;6(3):239-246. doi: 10.1016/j.ogla.2022.11.005. Epub 2022 Nov 24.

Abstract

Purpose: Portable perimetric testing could be useful for community-based glaucoma screening programs. Frequency-doubling technology (FDT) and the Moorfields motion displacement test (MDT) are portable perimeters that have shown promise as potential screening tools for glaucoma. This study's goal was to determine the diagnostic accuracy of FDT and MDT for visual field defects and glaucoma.

Design: Prospective, cross-sectional, diagnostic accuracy study.

Participants: A consecutive series of patients aged ≥ 50 years who presented to a glaucoma clinic in South India and had never undergone Humphrey field analyzer (HFA) visual field testing in the past.

Methods: Participants underwent 24-2 Swedish Interactive Thresholding Algorithm (SITA) Standard HFA perimetry, FDT perimetry, MDT perimetry, and iPad perimetry using visualFields Easy in random order. Ophthalmologist grades of HFA and optic nerve head photographs were used as reference standards for glaucoma and field defect presence. Receiver operating characteristic curves were constructed to assess the diagnostic accuracy of various parameters for each test.

Main outcome measures: Sensitivity, specificity, and area under the receiver operating characteristic curve (AUROC).

Results: Overall, 292 eyes from 173 participants were included, with 112 eyes classified as moderate or worse glaucoma. For moderate or worse glaucoma detection, the best parameter on FDT was mean deviation (MD) (AUROC, 0.84; 95% confidence interval [CI], 0.79-0.89) and the best parameter on MDT was global probability of true damage (GPTD) (AUROC, 0.87; 95% CI, 0.82-0.91). When specificity was set to 90%, the sensitivity for detection of moderate or worse glaucoma was 55% (95% CI, 39%-68%) for FDT MD and 62% (95% CI 52%-71%) for MDT GPTD.

Conclusions: Frequency-doubling technology and MDT perimetry had fair diagnostic accuracy for glaucoma detection when administered to naïve test takers in this South Indian population. Although not appropriate for use as a sole glaucoma screening test, these perimetric tests may be useful as ancillary tests.

Financial disclosure(s): Proprietary or commercial disclosure may be found after the references.

Keywords: Frequency-doubling technology, Glaucoma, Moorfields motion displacement test, Visual fields.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, N.I.H., Extramural

MeSH terms

  • Cross-Sectional Studies
  • Glaucoma* / diagnosis
  • Humans
  • Prospective Studies
  • Sensitivity and Specificity
  • Visual Field Tests
  • Visual Fields*