Tapered self-expandable metallic stent optimized for Eustachian tube morphology in a porcine ET model

Sci Rep. 2022 Nov 24;12(1):20290. doi: 10.1038/s41598-022-24615-6.

Abstract

Several investigations on the feasibility of stent placement into the Eustachian tube (ET) are being conducted. However, stents optimized for the anatomical structure of the ET have not yet been developed. In this study, the efficacy and safety of a self-expandable metallic stent (SEMS) optimized for porcine ET morphology was investigated. Silicone was injected into a cadaveric porcine ET to analyze the ET morphology. The three-dimensional-reconstructed porcine ET phantom images obtained after a computed tomography scan were measured to determine the dimensions of the porcine ET. The SEMS was designed as a tapered structure on the basis of the morphological findings of the porcine ET. The tapered SEMS (T-SEMS) and conventional SEMS (C-SEMS) were placed into the porcine ET to compare the safety and efficacy of the two types of SEMSs. Stent-induced tissue hyperplasia in the T-SEMS group was significantly lower than that in the C-SEMS group (p < 0.001). The T-SEMS optimized for the porcine ET was effective in maintaining stent patency. T-SEMS seems to be better than C-SEMS in suppressing stent-induced tissue hyperplasia, owing to the reduced stent-mediated mechanical injuries and maintaining ET patency.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Eustachian Tube* / diagnostic imaging
  • Eustachian Tube* / surgery
  • Hyperplasia
  • Retrospective Studies
  • Self Expandable Metallic Stents*
  • Stents
  • Swine